Measuring Method |
Laser Diffraction Method |
Measuring Range |
Basic |
Wet: 0.70 ~ 1,000um, Dry: 0.70 ~ 1,000um |
High |
Wet: 2.75 ~ 2,800um, Dry: 2.75 ~ 2,800um |
Standard |
Wet: 0.24 ~ 1,400um, Dry: 0.24 ~ 1,400um |
Special |
Wet: 0.08 ~ 1,400um, Dry: 0.24 ~ 1,400um |
Extended |
Wet: 0.02 ~ 2,000um, Dry: 0.24 ~ 2,000um |
Enhanced |
Wet: 0.02 ~ 2,800um, Dry: 0.24 ~ 2,800um |
Precision |
Spherical Glass Beads D50 = 642 micron, Precision as CV = 0.7%
Spherical Glass Beads D50 = 56 micron, Precision as CV = 1.0%
Spherical Latex Beads D50 = 0.40 micron, Precision as CV = 0.6% |
Lasers |
Wavelength 780nm, Power 3mW nominal |
Detection System |
Two fixed photo-electric detectors with logarithmically spaced
segments placed at correct angles for optimal scattered light detection. 0.02 to 163 degrees using 151detector segments |
Data Handling |
Volume, Number and Area distributions as well as percentile and other
summary data. Data is stored in ODBC format in encrypted Microsoft
Access Databases to ensure compatibility with external statistical software applications. Data integrity may be ensured using FDA 21 CFR Part 11 compliant security features including password protection, electronic signatures and assignable permissions. |
Typical Analysis Time |
10 to 30 seconds |
Electrical |
AC input: 90 - 132 VAC, 47 - 63 Hz, single phase
200 to 265 VAC, 47 ? 63 Hz, single phase |
Power Consumption |
25 VA maximum |
Environmental |
Temperature: 10 to 35 Degrees C.
Humidity: 90% RH, non condensing maximum
Storage Temperature: -10 to 50 Degrees C (Dry only)
Pollution: Degree 2 |
Physical Specifications |
Case Material; Steel and impact resistant plastic |
Finishes |
Exterior Surfaces finished with corrosion resistant paint or plating |
Dimensions |
14H x 22W x 18D in (360H x 560W x 460D mm) |
Weight |
60 lbs (27 kg) |
Dry Operation |
|
Eductor Air Supply: |
100 psi (689 kPa) maximum pressure
5 CFM at 50 psi (345 kPa) minimum flow rate
Free of dry contaminants, moisture and oil. |
Vacuum: |
No greater than 50 CFM |
|